 |
 |
 |
Overview
Importance of validation of results
As with the portable instruments, the importance of result validation during
measurements
must be stressed. The calibration certificates that were
enclosed
with
both HygroLab 2 and LabMASTER-aw were found to be of no value.
It is recommended that, at least initially, the instruments are validated using
calibration samples every day.
Below follows a summary of the three desktop instruments tested.
Click on instrument image to see measurement results diagram.
First a discussion about calibration samples and salt slurries must be given.
|
| Salt slurries vs. Decagon calibration liquids |
| |
As seen in the results from
AquaLab,
HygroLab and
LabMASTER-aw (SatSalt2),
saturated salt slurries (including Novasina calibration standards)
and
Decagon calibration solution "Dec LiCl 13.41m" with claimed aw = 0.250,
gave inconsistent results.
The inconsistency was in the order 0.01 aw.
A slighter inconsistency was possibly also seen for the sample "Dec LiCl 8.57m".
|
| Decagon AquaLab 3TE |
| |
The instrument showed very good reproducibility, also after
11 days.
The deviations for the Decagon calibrations samples
at aw = 0.75, 0.50 and 0.25
were extremely small.
The deviation at aw = 0.98 raises questions however,
as the results at
aw = 0.996 and 1.00 were extremely good.
A major question with this instrument, however,
seems to be
the
basis of instrument calibration.
|
 |
| |
Decagon uses their own calibration samples.
As seen above, the measurement results
from
at least one of these samples do not match results from saturated salt solutions.
Documentation showing
how the
calibration samples
are validated has
not
been found, despite requests from EminTech to Decagon.
As AquaLab 3TE obviously is calibrated using Decagon calibration samples,
it must be concluded
that
the factory calibration is not fully documented.
As shown in this link,
a crude correction
of the AquaLab results will give reasonable correlation with
literature
values for the salt slurries and with results from a recently
calibrated
LabMASTER-aw.
Once the calibration issues are cleared out, the impressing reproducibility and
easy
handling
of
the
instrument
makes AquaLab 3TE worth a strong recommendation.
Until then, the
instrument
is
still worth serious consideration for
high quality measurements of water activity.
|
|
HygroLab 2 with sample cell AwVC-DIO |
| |
Once HygroLab was calibrated, the instrument gave results
within specifications,
except at higher aw.
Calibration took at least 2 hours.
This instrument was lower specified than AquaLab
and
LabMASTER-aw,
but was also considerably less expensive. |
 |
| |
The possibility of analysing larger sample volumes was one of the factors
that
makes
this
instrument well worth consideration.
|
|
Novasina LabMASTER-aw |
| |
Once LabMASTER-aw was calibrated, measurements
using
the calibration
samples gave results according
to
specifications, except at higher aw.
Calibration took at least 3 hours.
The first analysis series of the salt slurries,
SatSalt2,
gave reasonable results
with one exception.
Analysis at higher aw still appeared uncertain. |
 |
| |
Repeating the measurements after 11 days, gave
results
that were
about
0.03
higher than previously,
i.e. the instrument required calibration again.
Provided calibration routines are followed, LabMASTER-aw may very well be used
when high quality results
from samples with aw less than 0.98 are required.
The good analysis temperature stability and the
possibility
to store various types of
analyses in the instrument are factors that add to the qualities of this instrument.
|